• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG INLEXA 3 HF-T DF4 IS-1 PROMRI; CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG INLEXA 3 HF-T DF4 IS-1 PROMRI; CRT-D Back to Search Results
Model Number 404700
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Event Description
An unexpected decrease of the remaining battery charge was observed approx.59 months after the implantation.The device was explanted.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.There was no indication of a material or manufacturing problem.Subsequently the device was interrogated.The interrogation could be properly performed and revealed the mos2 battery status.All electrical parameters were verified and showed a normal device behavior.Especially, the current consumption and the battery state of discharge were found normal and as expected.There was no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INLEXA 3 HF-T DF4 IS-1 PROMRI
Type of Device
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key17275290
MDR Text Key318562755
Report Number1028232-2023-03369
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Model Number404700
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/06/2023
Initial Date FDA Received07/06/2023
Supplement Dates Manufacturer Received09/08/2023
Supplement Dates FDA Received09/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-