Model Number 176630 |
Device Problems
Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 06/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
D10 concomitant product: egia45amt egia 45 artic med thick sulu (lot#: p3a0030) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, in a laparoscopically assisted distal gastrectomy, during the fourth firing for gastroduodenal anastomosis after duodenal and stomach resection, there was poor staple formation and the distal staple line was incomplete.The staple on the furthest tip of the cartridge was completely unformed.Due to the possibility of the crotch part of the staple line area not anastomosed due to staple non-formation, it was repaired and reinforced by ligation which resulted to tissue damage.There was also an issue with a clip applier used.The clip applier fired clips in succession by only squeeze the handle once, but there was no problem with clipping itself, so it was used as is.The operation time was extended by 30 minutes or more.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the instrument was received partially applied and the clip counter was not active.No visual abnormalities were observed with the instrument.Functionally, the instrument was cycled to load a clip, and when the handle was released partway through the cycle an unformed clip remained in the jaws.It was reported that the clip(s) were unintentionally ejected from the jaws.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the ratchet system was not functioning properly, but the failure could not be determined.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|