Model Number 71992-01 |
Device Problem
Failure to Fire (2610)
|
Patient Problem
Hypoglycemia (1912)
|
Event Date 06/18/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
Customer reported attempting to apply an adc device and being unsuccessful due to the inserter not firing, and as a result of being unable to monitor glucose levels, user reported feeling "sweaty and cold." as a result, the customer required third-party treatment of juice.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Sensor kit expiration date was determined to be 31 dec 2021.Medical event date of this issue was 1 jun 2023, which confirms the usage beyond the useful life of the device.Additional investigation activities are not required at this time as the device met specification when it was released and throughout its lifespan.If the product is returned, the case will be re-opened, and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
Customer reported attempting to apply an adc device and being unsuccessful due to the inserter not firing, and as a result of being unable to monitor glucose levels, user reported feeling "sweaty and cold." as a result, the customer required third-party treatment of juice.There was no report of death or permanent impairment associated with this event.
|
|
Search Alerts/Recalls
|