Catalog Number 3301-1 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
|
Event Description
|
It was reported that during final tightening, four closure tops stripped.There was a 30-minute delay and no impact on the patient.This is report two of four for this event.
|
|
Manufacturer Narrative
|
Corrections in d9 and h3.Additional information in h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the hex has deformation damage.Root cause: root cause was unable to be determined.This event could possibly be attributed to off-axis forces applied during closure top insertion or using a stripped driver.Dhr review: per dhr review, the part was likely conforming when it left highridge control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference reports 3012447612-2023-00210 through 3012447612-2023-00213.
|
|
Event Description
|
It was reported that during final tightening, four closure tops stripped.There was a 30-minute delay and no impact on the patient.This is report two of four for this event.
|
|