Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 06/13/2023
Event Type  Injury  
Event Description
It was reported that a patient underwent left knee revision surgery due to unknown reason approximately six (6) years after initial surgery.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).D10 - oxford pks cemented tibia, alpha, standard - lm size b, item# 154720, lot# 777150; oxford pks anatomic arcom meniscal brg - small left size 5, item# 159542, lot# 735940.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2023-00231, 3002806535-2023-00232.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.No products were returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Devices are used for treatment.Radiographs were provided and reviewed by a radiologist.A medial unicompartmental arthroplasty is present.There is abnormal position of the tibial implant with oblique orientation on the lateral view and probable posterior implant subsidence.Alignment is overall maintained.Correlation with earlier images would be of value.Bone quality is osteopenic.No other abnormalities are identified.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF TWIN-PEG CMNTD FEM SM PMA
Type of Device
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17286100
MDR Text Key318648928
Report Number3002806535-2023-00233
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279438228
UDI-Public(01)05019279438228(17)270126(10)913310
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number161468
Device Lot Number913310
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
-
-