MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; NO MATCH Back to Search Results
Model Number PM3242
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the lead was unable to be removed from the pacemaker's header.The lead was eventually disconnected after removing the set screw.The pacemaker was explanted and replaced.The patient experienced no consequences.
 
Manufacturer Narrative
The reported event of failure to remove lead from the header was confirmed.Analysis revealed the device has the original scalable brady pacemaker (sbp) header design, which is known to have lead insertion and removal difficulties.The original sbp header required extra force to insert and remove leads from the connector ports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17286191
MDR Text Key318654792
Report Number2017865-2023-25376
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507912
UDI-Public05414734507912
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2015
Device Model NumberPM3242
Device Catalogue NumberPM3242
Device Lot Number4630785
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/16/2023
Initial Date FDA Received07/10/2023
Supplement Dates Manufacturer Received08/11/2023
Supplement Dates FDA Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-