Model Number FINETUNER ECHO [FINETUNER 2] |
Device Problems
No Device Output (1435); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Event Description
|
A finetuner echo was returned to service and repair.Finetuner echo can no longer be unlocked despite new battery.
|
|
Event Description
|
A finetuner echo was returned to service and repair.Finetuner echo can no longer be unlocked despite new battery.
|
|
Manufacturer Narrative
|
Conclusion: according the information received, the fine tuner echo could no longer be unlocked.The investigation revealed two loosened components, on of which was the tantalum capacitor, which was clearly the reason for the reported issue.Observed device malfunction was most likely caused by mechanical stress, which may have been inadvertently exerted during final assembly and/or device usage.This is a final report.
|
|