BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE
|
Back to Search Results |
|
Catalog Number 368836 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that while using the bd vacutainer® eclipse¿ signal¿ blood collection needle with integrated holder that there was a clogged or blocked cannula.The following information was provided by the initial reporter: in this department, many assessments have to be carried out, with up to 9 tubes taken in most cases.The nurses told us that, with eclipse signal needles, after 3 tubes, there was no reflux to fill the tubes, forcing them to prick the patient again.As a result, they fall back on the wings.
|
|
Manufacturer Narrative
|
H.6 investigation summary: material #: 368836, lot/batch #: 2305497, bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by functional testing, each used to draw 6 vacutainer tubes with water, and no issues were observed relating to clog as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode clog.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|
|
|