Model Number N/A |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the unit ran slow.There is no harm or delay reported.The event timing is outside of surgery.No adverse events were reported as a result of this malfunction.Due diligence is in progress and there is no additional information till date.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined the spring seal was stretched and bearings replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|
Manufacturer Narrative
|
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
The event timing was while getting ready for use and another device was used to finish the surgery.
|
|
Search Alerts/Recalls
|