It was reported that on (b)(6)2023, a 30-25mm amplatzer talisman pfo occluder was implanted.There was a deflection from the septum, and the defect was measured as 15mm.The sizing of the device was determined via transesophageal echocardiogram (tee).On (b)(6)2023, the patient presented to the clinic for a follow-up appointment.The device was not able to be found on imaging, and the pfo was observed to be open.The location of the occluder was not able to be identified.The patient was transferred and admitted to another hospital.X-ray and computed tomography (ct) scan was conducted.The device was located in the right pulmonary vein causing a partial obstruction.On (b)(6)2023, the device was explanted via surgical intervention.The patient was reported as stable.
|
An event of an occluder not being found on imaging during a follow up appointment was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|