It was reported that prior to a cryo ablation procedure, when the integrated dilator/needle was inserted into the sheath it would not lock into the hub of the sheath.The physician tried to push with force, however the dilator would not lock.The dilator was replaced which resolved the issue.There was no patient involvement.
|
Product event summary: the 4fc12 sheath with lot number 0011563069 was returned and analyzed.Visual inspection of the shaft area was performed.The inspection identified a shaft kink/twist at approximately 25.5 inches from the tip.The dilator was not returned with the sheath.The test dilator was inserted into the sheath with friction and difficulty.It was snap locked in the sheath properly.In conclusion, the sheath failed the returned product inspection due to the shaft kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|