Model Number 72081-01 |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
Loss of consciousness (2418)
|
Event Date 06/22/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A bent sensor tip was reported with the adc device and customer was therefore unable to apply the device and monitor glucose levels.As a result, the customer experienced a loss of consciousness, redness at the sensor site and was able to self-treat with unspecified oral medication.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.Visual inspection was performed on the returned sensor patch, no issues were observed.No issues were observed with the sensor adhesive.A watermark at the base of the sensor tail was observed, this is an indication that the sensor was properly inserted.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A bent sensor tip was reported with the adc device and customer was therefore unable to apply the device and monitor glucose levels.As a result, the customer experienced a loss of consciousness, redness at the sensor site and was able to self-treat with unspecified oral medication.There was no report of death or permanent impairment associated with this event.
|
|