SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Model Number C37101307 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: b3: month and year are known, date is unknown.D4: udi number, g5 - unknown, no information to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that during the pre-use check, the customer noticed no l-connector came with the product set.No patient involvement.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer and h6.Evaluation codes: updated.Four photos and one device was received without it?s original package.Visual inspection noted the product was missing the "elbow" component and embedded resin was found on the corrugated tube.The complaint was confirmed.No other analysis was performed.Root cause was during the elbow connector by operator omission.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.Actions taken; increase "aql" inspection before the material is packed.This issue will be monitored, and further actions taken accordingly.
|
|