• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number D134805
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a foreign material inside the packaging was discovered.It was reported that the catheter has a "dirt particle" inside the sterile sleeve.Caller stated they never opened the catheter and decided to use a different catheter for the procedure.Issue was noticed before use so it was never opened.The debris was described as small black pieces that were lose.
 
Manufacturer Narrative
The product has not returned for analysis, however, a picture was provided by the customer.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
 
Manufacturer Narrative
On 9-aug-2023, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Manufacturer Narrative
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a foreign material inside the packaging was discovered.It was reported that the catheter has a "dirt particle" inside the sterile sleeve.Caller stated they never opened the catheter and decided to use a different catheter for the procedure.Issue was noticed before use so it was never opened.The debris was described as small black pieces that were lose.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection and fourier transformed infrared spectroscopy (ft-ir) test of the returned device were performed following bwi procedures.Visual analysis revealed that a particle was observed inside the pouch of the catheter (a small black debris piece).A fourier transformed infrared spectroscopy (ft-ir) was performed and based on the results obtained it can be concluded that black particle material match on all infrared absorption bands with rocker arm reference spectrum, this result suggest that the rocker arm component can be identified as the source of origin of the black material.Additionally, according to pictures provided by the customer, it was observed a black unknown material inside a plastic bag.Due to the conductions found, a manufacturing investigation was performed to determine the root cause of this issue.A manufacturing investigation was performed and it was determined that this issue is related to the manufacturing process since the small black debris are part of the rocker arm component.An awareness session was provided to the involved personal.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17302104
MDR Text Key318938961
Report Number2029046-2023-01476
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD134805
Device Catalogue NumberD134805
Device Lot Number31056660L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/13/2023
Initial Date FDA Received07/11/2023
Supplement Dates Manufacturer Received08/09/2023
10/06/2023
Supplement Dates FDA Received08/14/2023
10/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
THMCL SMTCH SF BID, TC, D-F
-
-