Model Number D134805 |
Device Problem
Contamination /Decontamination Problem (2895)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a foreign material inside the packaging was discovered.It was reported that the catheter has a "dirt particle" inside the sterile sleeve.Caller stated they never opened the catheter and decided to use a different catheter for the procedure.Issue was noticed before use so it was never opened.The debris was described as small black pieces that were lose.
|
|
Manufacturer Narrative
|
The product has not returned for analysis, however, a picture was provided by the customer.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
|
Manufacturer Narrative
|
On 9-aug-2023, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
|
Manufacturer Narrative
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and a foreign material inside the packaging was discovered.It was reported that the catheter has a "dirt particle" inside the sterile sleeve.Caller stated they never opened the catheter and decided to use a different catheter for the procedure.Issue was noticed before use so it was never opened.The debris was described as small black pieces that were lose.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection and fourier transformed infrared spectroscopy (ft-ir) test of the returned device were performed following bwi procedures.Visual analysis revealed that a particle was observed inside the pouch of the catheter (a small black debris piece).A fourier transformed infrared spectroscopy (ft-ir) was performed and based on the results obtained it can be concluded that black particle material match on all infrared absorption bands with rocker arm reference spectrum, this result suggest that the rocker arm component can be identified as the source of origin of the black material.Additionally, according to pictures provided by the customer, it was observed a black unknown material inside a plastic bag.Due to the conductions found, a manufacturing investigation was performed to determine the root cause of this issue.A manufacturing investigation was performed and it was determined that this issue is related to the manufacturing process since the small black debris are part of the rocker arm component.An awareness session was provided to the involved personal.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
|