• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 06/13/2023
Event Type  Injury  
Manufacturer Narrative
At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the libre sensor and sensor kit were reviewed, and the dhrs showed the libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date the incident occurred is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A webform complaint was received in which it was reported a customer experienced a bleeding issue with the adc device.The customer was given unspecified treatment by a third-party.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor applicator (b)(6) has been returned and investigated.Visual inspection has been performed on the applicator and no issues were observed.Applicator had fired correctly.The applicator was pried open to inspect the sharp; no issues was observed with the sharp.Sensor pack (b)(6) has been returned and investigated.Visual inspection has been performed on the sensor pack and observed damaged transition features.Lid was completely peeled off.Unable to perform further investigation due to no product returned.Issue close to no product returned.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A webform complaint was received in which it was reported a customer experienced a bleeding issue with the adc device.The customer was given unspecified treatment by a third-party.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key17302536
MDR Text Key318824869
Report Number2954323-2023-29794
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 06/13/2023
Initial Date FDA Received07/11/2023
Supplement Dates Manufacturer Received07/15/2023
Supplement Dates FDA Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-