The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Testing measurements were within normal limits excepting shock range which was below specification.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|