|
Model Number 10672 |
Device Problems
Deflation Problem (1149); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that deflation issue occurred.The 70% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 16 x 3.50 promus elite drug-eluting stent was advanced for treatment.However, during the procedure, the stent balloon had a leak and was not inflating and deflating properly.The device was removed, and the procedure was completed with a 16 x 3.00 promus elite drug-eluting stent with timi 3 flow.No patient complications were reported.
|
|
Manufacturer Narrative
|
The promus elite ous mr 16 x 3.50mm was returned for analysis.Visual and tactile examination revealed no kinks or damage to the hypotube shaft profile or shaft polymer extrusion.Microscopic inspection of the shaft polymer extrusion profile revealed a pinhole lead in the distal extrusion and no damage to the inner/wire lumen at the site of the pinhole leak.The leak appeared to be a puncture in the outer lumen.No issues were identified with the lumen which could have contributed to the leak.A microscopic examination of the stent found no damage; the stent was securely crimped to the balloon.A microscopic examination of the top section found on damage.Leakage testing revealed a pinhole leak located 5.4cm from the tip of the device on the polymer extrusion.As a result of the leak, the balloon could not inflate properly and as a result just the proximal stent struts started to partially deploy.
|
|
Event Description
|
It was reported that deflation issue occurred.The 70% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 16 x 3.50 promus elite drug-eluting stent was advanced for treatment.However, during the procedure, the stent balloon had a leak and was not inflating and deflating properly.The device was removed, and the procedure was completed with a 16 x 3.00 promus elite drug-eluting stent with timi 3 flow.No patient complications were reported.
|
|
Search Alerts/Recalls
|
|
|