OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTOSCOPE
|
Back to Search Results |
|
Model Number A42011A |
Device Problems
Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2023 |
Event Type
malfunction
|
Event Description
|
The customer returned the resection sheath, to olympus repair center for evaluation of a reported loose ceramic tip that was found during reprocessing.There was no delay, no procedure involved and there were no reports of patient harm.
|
|
Manufacturer Narrative
|
The subject device was returned to olympus for evaluation.The device evaluation found the ceramic tip was detached.Furthermore, the following additional issues were identified during inspection: the serial numbers are worn, and the sealing ring is damaged.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, the loosening of the ceramic tip was caused by use of a third-party adhesive.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
|
|