MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTOSCOPE Back to Search Results
Model Number A42011A
Device Problems Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Event Description
The customer returned the resection sheath, to olympus repair center for evaluation of a reported loose ceramic tip that was found during reprocessing.There was no delay, no procedure involved and there were no reports of patient harm.
 
Manufacturer Narrative
The subject device was returned to olympus for evaluation.The device evaluation found the ceramic tip was detached.Furthermore, the following additional issues were identified during inspection: the serial numbers are worn, and the sealing ring is damaged.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, the loosening of the ceramic tip was caused by use of a third-party adhesive.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS
Type of Device
RESECTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17313315
MDR Text Key319103693
Report Number9610773-2023-01864
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761023658
UDI-Public04042761023658
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA42011A
Device Catalogue NumberA42011A
Device Lot Number195W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-