ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 15" (38 CM) APPX 4.7 ML, BIFUSE ADD-ON SET W/2 BAG SPIKES, 2 CLAMPS (RED, BLUE),; STOPCOCK, I.V. SET
|
Back to Search Results |
|
Catalog Number CH3139 |
Device Problem
Particulates (1451)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/29/2023 |
Event Type
malfunction
|
Event Description
|
The incident involved a 15" (38 cm) appx 4.7 ml, bifuse add-on set w/2 bag spikes, 2 clamps (red, blue), plug adapter on an unknown date in june 2023.The reporter stated that after the nurses prepare their lines by attaching the spike to saline, then attaching the iv line, plastic is noted in the iv line chamber when priming with saline.The sets were not reprocessed or re-sterilized prior to use.There was no patient involvement, no delay in therapy, and no one was harmed as a result of the reported event.This is the second of two reports.
|
|
Manufacturer Narrative
|
The device is available to be returned for evaluation; however, it has not yet been received.
|
|
Manufacturer Narrative
|
A used sample #ch3139 connected into a 0.9% sodium chloride 500ml bag and an unknown, extension set w/ drip chamber was returned for evaluation.As received inside of the dry chamber, a small debris was found floating.Inside the dry spike adaptor, it was observed an excess of plastic.The set was tested using a filtration system, and a small plastic debris was found belonging from the spike adaptor.Complaint of particulate matter can be confirmed based in the physical sample returned by customer.The probable cause of the excess of plastic inside the dry spike adaptor was due an error during molding process from manufacturer¿s (ensenada) supplier.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
|
Manufacturer Narrative
|
The customer's complaint of particulate matter can be confirmed based in the physical sample returned by customer.No assignable cause related to icu medical product manufacturing or design was identified.Reported condition of the returned device was replicated.The shape of this drip chamber used to access the icu medical dry spike is considered the most relevant factor to create these particulates.The probable cause was that the drip chamber tip got broken inside the dry spike adaptor due to unintentional bending force applied during use.Updated information can be found in g1.D9 date returned to mfg: 08aug2023.
|
|