• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 15" (38 CM) APPX 4.7 ML, BIFUSE ADD-ON SET W/2 BAG SPIKES, 2 CLAMPS (RED, BLUE),; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 15" (38 CM) APPX 4.7 ML, BIFUSE ADD-ON SET W/2 BAG SPIKES, 2 CLAMPS (RED, BLUE),; STOPCOCK, I.V. SET Back to Search Results
Catalog Number CH3139
Device Problem Particulates (1451)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2023
Event Type  malfunction  
Event Description
The incident involved a 15" (38 cm) appx 4.7 ml, bifuse add-on set w/2 bag spikes, 2 clamps (red, blue), plug adapter on an unknown date in june 2023.The reporter stated that after the nurses prepare their lines by attaching the spike to saline, then attaching the iv line, plastic is noted in the iv line chamber when priming with saline.The sets were not reprocessed or re-sterilized prior to use.There was no patient involvement, no delay in therapy, and no one was harmed as a result of the reported event.This is the second of two reports.
 
Manufacturer Narrative
The device is available to be returned for evaluation; however, it has not yet been received.
 
Manufacturer Narrative
A used sample #ch3139 connected into a 0.9% sodium chloride 500ml bag and an unknown, extension set w/ drip chamber was returned for evaluation.As received inside of the dry chamber, a small debris was found floating.Inside the dry spike adaptor, it was observed an excess of plastic.The set was tested using a filtration system, and a small plastic debris was found belonging from the spike adaptor.Complaint of particulate matter can be confirmed based in the physical sample returned by customer.The probable cause of the excess of plastic inside the dry spike adaptor was due an error during molding process from manufacturer¿s (ensenada) supplier.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
 
Manufacturer Narrative
The customer's complaint of particulate matter can be confirmed based in the physical sample returned by customer.No assignable cause related to icu medical product manufacturing or design was identified.Reported condition of the returned device was replicated.The shape of this drip chamber used to access the icu medical dry spike is considered the most relevant factor to create these particulates.The probable cause was that the drip chamber tip got broken inside the dry spike adaptor due to unintentional bending force applied during use.Updated information can be found in g1.D9 date returned to mfg: 08aug2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
15" (38 CM) APPX 4.7 ML, BIFUSE ADD-ON SET W/2 BAG SPIKES, 2 CLAMPS (RED, BLUE),
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key17316462
MDR Text Key318963957
Report Number9617594-2023-00370
Device Sequence Number1
Product Code FMG
UDI-Device Identifier00840619024918
UDI-Public(01)00840619024918(17)271201(10)13408928
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K964435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCH3139
Device Lot Number13408928
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2023
Initial Date FDA Received07/13/2023
Supplement Dates Manufacturer Received09/25/2023
04/04/2024
Supplement Dates FDA Received10/02/2023
05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-