• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S FOLYSIL SILICONE CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S FOLYSIL SILICONE CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL Back to Search Results
Catalog Number AA6C20
Device Problem Nonstandard Device (1420)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Event Description
According to the available information, the catheter did not have irrigation eyelets.
 
Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
 
Manufacturer Narrative
After receiving this complaint, we searched for other complaint and we didn¿t find other complaint on the lot n° 8679898.Checking the quality databases revealed no anomaly in connection with the described defect.On 11 october we received subcontractor investigation.After analysis of the picture the step of cutting was not performed.A resensibilisation of operator was done in order to avoid this defect.Occured again.
 
Event Description
We have been informed about a defect product, irrigation eye missing on a folysil long term product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL SILICONE CATHETER
Type of Device
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
CMF-SARLAT
9 avenue edmond rostand
sarlat-la-caneda
FR  
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17317575
MDR Text Key319748695
Report Number9610711-2023-00139
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
PMA/PMN Number
K013174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAA6C20
Device Lot Number8679898_AA6C201002
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-