Catalog Number AA6C20 |
Device Problem
Nonstandard Device (1420)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2023 |
Event Type
malfunction
|
Event Description
|
According to the available information, the catheter did not have irrigation eyelets.
|
|
Manufacturer Narrative
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
|
Manufacturer Narrative
|
After receiving this complaint, we searched for other complaint and we didn¿t find other complaint on the lot n° 8679898.Checking the quality databases revealed no anomaly in connection with the described defect.On 11 october we received subcontractor investigation.After analysis of the picture the step of cutting was not performed.A resensibilisation of operator was done in order to avoid this defect.Occured again.
|
|
Event Description
|
We have been informed about a defect product, irrigation eye missing on a folysil long term product.
|
|
Search Alerts/Recalls
|