• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VAN PS OPEN INTL FEM-LT 70; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VAN PS OPEN INTL FEM-LT 70; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Naturally Worn (2988)
Patient Problem Metal Related Pathology (4530)
Event Date 04/20/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).G2: spain.The product will not be returned to zimmer biomet for investigation, as the device location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the patient underwent a knee revision due to short-term prosthetic deterioration, with increased wear of the implant and increased metals in the joint (metalosis) and systemic blood (cobalt).Original implant date is not known.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  h6-component code- suggested code: mechanical(04)- femur.Visual examination of the provided pictures identified that the femoral implant is implanted, has some surface scratches and is covered in foreign material.As the implant was not returned no other definitive statements can be made.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Insufficient information provided to perform a compatibility check.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.Complaint is confirmed from provided pictures.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAN PS OPEN INTL FEM-LT 70
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17321262
MDR Text Key319023756
Report Number0001825034-2023-01629
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00880304270923
UDI-Public(01)00880304270923(17)250210(10)171420
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K171054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/03/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received07/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number183132
Device Lot Number171420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-