Trackwise #(b)(4).Corrected section: h6-medical device ¿ problem code- corrected to "1158".The device was returned to the factory for evaluation on 07/19/2023.An investigation was conducted on 08/01/2023.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the device.A mechanical evaluation was conducted.The positioner was attached and locked onto a reference retractor blade by sliding the stabilizer base onto the rail and locking the lever.The flexlink arm was able to be adjusted and positioned, and then locked into place by turning the knob clockwise with no physical or visual difficulties, with the locking lever in both the locked and unlocked positions.Based on the returned condition of the device as well as the evaluation results, the reported failure ¿positioning problem" was not confirmed.The lot # 3000288696 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|