H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 04/2024) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was not returned for evaluation.No photos were provided which leads to inconclusive results.Based on the available information and as the sample was not returned, the investigation is closed with inconclusive results.A definite root cause of the reported incident can not be identified.Labeling review: in reviewing the relevant labeling it was found that the instructions for use sufficiently address the potential risks.With regards to warnings, the instructions for use states "visually inspect the lifestar vascular stent system to verify that the device has not been damaged due to shipping or improper storage.Do not use damaged equipment".Regarding accessories, the instructions for use states "the 6f delivery system requires a minimum 8f guiding catheter, or a minimum 6f introducer sheath.Via the femoral route, insert a 0.035 inch (0.89 mm) guidewire under fluoroscopic guidance through the appropriate introducer sheath or guiding catheter and pass the lesion".The packaging pictograms indicate an introducer size of 6f and a 0.035" guidewire.With regards to the procedure, the instructions for use states "pre-dilatation of the stricture is recommended.Selection of an appropriately sized balloon dilatation catheter is left to the discretion of the treating physician".H10: d4 (expiry date: 04/2024), g3 h11: h6 (result, conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|