Model Number 4FC12 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id afapro28; product type: balloon catheter; product id 990063-020; product type: mapping catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, the balloon catheter was inserted into the sheath and air continued to be drawn when the reverse blood was drawn.The balloon catheter position was shifted which did not resolve the issue.The sheath was replaced which resolved the issue.The procedure proceeded.The esophageal temperature decreased while ablating the left inferior pulmonary vein (lipv) and a double stop was performed.Immediately following the double stop, a system notice was received indicating that the safety system detected fluid in the catheter and stopped the injection.The procedure continued without replacing the balloon catheter.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the 4fc12 sheath with lot number 0011685271 was returned and analyzed.Visual inspection before functional testing and dissection was performed on the shaft, handle.No anomaly was identified during the external visual inspection.All the handle, shaft and sideport were intact with no apparent issue.The steering mechanism and deflection test were performed.The shaft is bending as initially intended and is bending in the plane.There is no difficulty, no friction, or any noise in the steering mechanism.The performance test with sentinel blackbelt leak tester was performed.All performance tests were in the acceptable range.The shaft, side tube, and valve were all leak-tight with no apparent issue.In conclusion, the reported air ingress could not be confirmed through testing.The sheath did not fail the returned product inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|