Model Number BD-400P-2080 |
Device Problem
Deflation Problem (1149)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is not expected to be returned for evaluation.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.H3 other text : device not returned.
|
|
Event Description
|
The customer reported to olympus that when using an ez dilate balloon dilator (fw) (b)(6) 2020, the device did not deflate all the way so it could not be pulled out through the scope.The user cut the balloon off the end to pull the catheter through.There was no patient harm associated with the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The investigation concluded that this nonconformance observed in this complaint was categorized as a process defect.An exhaustive revision of the process was completed, founding that occlusion partial or total was generated in the proximal bonding station due to misalignment of the d-mandrel or mandrel during the preparation for bonding process.The operator has not enough visibility to ensure the correct position of the d-mandrel or mandrel, causing melting material smearing occluding partial or total the diameter of the extrude.Causing two events, the first is the diameter is completely occluded which would prevent the balloon from inflating.And the second event would be that the diameter is partially occluded which would cause it to take longer to inflate or deflate.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|