• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. EZDILATE BALLOON DILATOR (FW) 18-19-20; BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. EZDILATE BALLOON DILATOR (FW) 18-19-20; BALLOON DILATATION CATHETER Back to Search Results
Model Number BD-400P-2080
Device Problem Deflation Problem (1149)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2023
Event Type  malfunction  
Manufacturer Narrative
The device is not expected to be returned for evaluation.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.H3 other text : device not returned.
 
Event Description
The customer reported to olympus that when using an ez dilate balloon dilator (fw) (b)(6) 2020, the device did not deflate all the way so it could not be pulled out through the scope.The user cut the balloon off the end to pull the catheter through.There was no patient harm associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The investigation concluded that this nonconformance observed in this complaint was categorized as a process defect.An exhaustive revision of the process was completed, founding that occlusion partial or total was generated in the proximal bonding station due to misalignment of the d-mandrel or mandrel during the preparation for bonding process.The operator has not enough visibility to ensure the correct position of the d-mandrel or mandrel, causing melting material smearing occluding partial or total the diameter of the extrude.Causing two events, the first is the diameter is completely occluded which would prevent the balloon from inflating.And the second event would be that the diameter is partially occluded which would cause it to take longer to inflate or deflate.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZDILATE BALLOON DILATOR (FW) 18-19-20
Type of Device
BALLOON DILATATION CATHETER
Manufacturer (Section D)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer (Section G)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17345797
MDR Text Key320219063
Report Number3003790304-2023-00280
Device Sequence Number1
Product Code KNQ
UDI-Device Identifier00821925033214
UDI-Public00821925033214
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBD-400P-2080
Device Lot Number383361
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-