ZIMMER BIOMET, INC. UNKNOWN GLENOID; SHOULDER, PROSTHESIS Back to Search Results
Catalog Number UNKNOWN
Device Problems Insufficient Information (3190); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Osteopenia/ Osteoporosis (2651); Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
(b)(4).It is unknown if product will be returning to zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported patient has been indicated for revision due to unknown reasons.No revision has been reported to date.Attempts to obtain additional information have been made; however, no more is available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: total shoulder arthroplasty with possible aseptic loosening of the glenoid component also with possible loosening of the humeral stem.Overall fit and alignment of the implant is appropriate.Radiolucency along the humeral stem and diffusely throughout the glenoid component at the bone cement interface is present.Osteopenia.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.The reported event is confirmed by mmi review.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that a patient underwent a left shoulder arthroplasty on an unknown date approximately 12 years ago.Subsequently, the patient is being considered for a revision on an unknown day due to glenoid aseptic loosening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN GLENOID
Type of Device
SHOULDER, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17345839
MDR Text Key319290518
Report Number0001825034-2023-01699
Device Sequence Number1
Product Code PHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/22/2023
Initial Date FDA Received07/18/2023
Supplement Dates Manufacturer Received08/14/2023
Supplement Dates FDA Received08/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN HEAD.; UNKNOWN STEM.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
-
-