• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE CORE 10-INCH MONITOR; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE CORE 10-INCH MONITOR; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0376
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2023
Event Type  malfunction  
Manufacturer Narrative
The customer declined the option to have their glidescope core 10-inch monitor returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Follow-up communication received from the customer indicated that they performed troubleshooting at their facility and reported that the system was working fine.No information was provided regarding the customer's resolution or the cable and/or laryngoscope used with the monitor during the reported event.Review of complaint history for the reported serial number " (b)(6)" did not identify any previous complaints reported to verathon.Trending analysis for the glidescope core monitors does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is currently not required at this time.Verathon will continue to monitor for any ongoing trends.
 
Event Description
A customer reported that during a patient procedure, using a glidescope core 10-inch monitor, the screen was freezing.No delay during the procedure, use of a backup device, or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE 10-INCH MONITOR
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key17347298
MDR Text Key319589627
Report Number9615393-2023-00137
Device Sequence Number1
Product Code CCW
UDI-Device Identifier00879123006509
UDI-Public010087912300650911201208
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0376
Device Catalogue Number0270-0996
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2023
Initial Date FDA Received07/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-