Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted minor cosmetic damage.Internal inspection found no apparent anomalies.Ac power was connected and unit was powered up.Rem (real ear measurement) testing was performed by both the service manual and by the eds (ehlers-danlos syndromes).Rem calibration was performed through the calibration menu.Rem testing was performed again and passed both the service manual and eds testing.Functional testing found that the unit failed both rem tests by remaining green and allowing the device to activate beyond the high threshold due to a rem calibration issue.The issue was resolved by calibrating the rem issue to address the condition.It was reported that the device had an unknown rem issue and it showed out of specification values or results.The reported issues were confirmed.The most likely cause was calibration malfunction.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|