The involved esu was inspected/tested.The unit was found to be functioning as intended.The evaluation included an electrical safety check, a functional check of each of the equipment's features and a power output check.The generator was/is within specifications and all features were/are functioning properly.In addition, no anomalies were found in the device history record (dhr) of the involved device.In conclusion, no erbe equipment problem was found that would have caused or contributed to the incident.Based upon the limited information provided, it appears that during activation, some of the current followed the path of least resistance to another electrically conductive device (e.G., a retractor) used in the procedure.Most likely, the current from the generator was high to overcome the resistance given by the coating of the needle electrode.This resulted in stray current to the retractor's hook which then burned a thin-skinned area of the patient (at the margin of the incision site).However, no conclusive determination could be made as to the cause of the event.No trends have been identified and erbe usa, inc.Is now closing the file on this event.
|
It was reported that a patient incident occurred with the electrosurgical unit (esu/generator) during an orchidopexy (i.E., a surgical procedure to free an undescended testicle and implanting it into the scrotum.).The esu was used with an insulated coated needle (note: no information was provided regarding the manufacturer, part number, etc.Of the instrument used.).Specific information involving the settings of the esu and other devices used was not provided.During the procedure, a small burn occurred at the edge of the surgical site.Therefore, the necrosis was removed before closing the incision.No details were given on the appearance, size, or location of the burn.
|