|
Model Number 311.431 |
Device Problem
Contamination /Decontamination Problem (2895)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a synthes employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2023 prior to a removal surgery, the screwdriver handle in question was found with a viscous liquid adhered to its surface.The nurse determined that it was not appropriate for use and the surgery was completed successfully using a substitute item.After the procedure, the handle was inspected and it was found that the cloth covering it was discolored.There was no surgical delay.No further information is available.This report is for a large handle with quick coupling.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9 h3, h6: part# 311.431.Lot # 9167720.Manufacturing site: werk bettlach.Supplier: na.Release to warehouse date: 10 dec 2014.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the handle w/quick-coupl presents light signs of fading of the brown color of the phenolic handle.These components can present this issue after several sterilization/cleaning cycles.The pouch in which the device was decontaminated and received prior to investigation presents light evidence of brown staining.The attached letter for sustaining engineering documents that the phenolic material, not this specific device, has been tested through numerous reprocessing cycles and slight color changes were noted but no signs of severe degradation such as bleeding or erosion were found.The returned device did not have evidence of brown staining or material leaching from the handle so the reported condition cannot be confirmed for the returned device.The returned instrument is over 8 years old and the slight fading noted on the phenolic handle would be expected for a device that has gone through numerous reprocessing cycles.A dimensional inspection for the handle w/quick-coupl was not performed since it is not applicable to the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the handle w/quick-coupl would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.G1 manufacturing site name and address.
|
|
Search Alerts/Recalls
|
|
|