• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7106-24
Device Problem Fluid/Blood Leak (1250)
Patient Problem Pulmonary Hypertension (4460)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported that the customer received a spike in complaints from patients related to leaky cadd extension set with male/male luer, 0.2-micron air-eliminating filter, integral anti-siphon valve, and clamp.Customer asked if there was a known defect with this tubing.Patients were asked to return their leaky extension sets to them for further evaluation.Received from accredo a report with malfunction from (b)(6) 2023, where 67 patients were affected.About injuries no major issues were reported, some patients mentioned exhibiting pulmonary arterial hypertension symptoms.
 
Manufacturer Narrative
Other, other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17350644
MDR Text Key319656934
Report Number3012307300-2023-07234
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586023279
UDI-Public10610586023279
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number21-7106-24
Device Catalogue Number21-7106-24
Device Lot Number4309417
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/10/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-