• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510880
Device Problems Use of Device Problem (1670); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2023
Event Type  Injury  
Manufacturer Narrative
Block h6: imdrf device code a150301 captures the reportable event of tip failure to separate.Imdrf impact code f05 captures the event that the patient was transferred to another facility for ercp.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) stone removal procedure performed on (b)(6) 2023.During the procedure, a trapezoid rx basket was used to attempt to remove a larger stone.However, the basket failed to crush the stone and the tip failed to separate.The duct was dilated to assist with pulling the basket and stone out of the duct.But the stone failed to be removed from the basket.The procedure was aborted and the patient was transferred to another facility for ercp with spyglass/ehl or laser to remove the stone.No patient complications were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carol morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17358309
MDR Text Key319373323
Report Number3005099803-2023-03597
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296393
UDI-Public08714729296393
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2024
Device Model NumberM00510880
Device Catalogue Number1088
Device Lot Number0031297553
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-