• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR LTD NUCLEUS CI24RE (CA) COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LTD NUCLEUS CI24RE (CA) COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI24RE (CA)
Device Problems Biofilm coating in Device (1062); Circuit Failure (1089); Electrical Shorting (2926); Expulsion (2933)
Patient Problem Skin Infection (4544)
Event Date 06/08/2023
Event Type  Injury  
Manufacturer Narrative
This report is submitted on july 20, 2023.
 
Event Description
Per the clinic, the patient experienced an extrusion of r/s and infection.On (b)(6) 2023, the patient underwent revision surgery to reposition the device.It was noted that a biofilm layer was observed on the r/s plate.Intra-operative tests were done and showed multiple open and short circuits, thus the device was explanted during the same surgery.The patient was re-implanted with a new cochlear device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI24RE (CA) COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
roberta loveday
1 university avenue
macquarie university, 2109
AS   2109
MDR Report Key17358545
MDR Text Key319371124
Report Number6000034-2023-02289
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502025690
UDI-Public(01)09321502025690(11)191112(17)211111
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/11/2021
Device Model NumberCI24RE (CA)
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age5 YR
Patient SexMale
-
-