ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Corroded (1131); Unstable (1667); Calibration Problem (2890); Naturally Worn (2988); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/26/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during pm on (b)(6) 2023 the unit had a control bar issue, motor speed unstable, calibration issue, motor issue, worn screws and corroded motor.There was no patient involvement.Due diligence is complete and there is no additional info.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete g2: united kingdom.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined the motor speeds were unstable (signs of corrosion), the screws were worn, the control bar was out of position, and the device was out of calibration at the 0 reading.The motor and screws were replaced, the control bar was repositioned, and the device was recalibrated and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|