The web delivery system pusher, four web detachment controllers, and microcatheter were all returned.The evaluation for all the returned devices is documented below.Device evaluation: the investigation of the returned web system found the implant separated from the delivery system, the hypotube kinked, the proximal connector kinked, and the heater coil windings stretched.The implant was not returned for evaluation.The device failed continuity and resistance testing due to the damaged connector and heater coil windings.However, the heater coil pet was found melted, indicating that the device was activated using a detachment controller during the procedure; therefore, the damage to the heater coil windings likely occurred post-activation.The stretched heater coil windings are an indication that the tether could have been caught in between the coil windings and caused the heater coil to stretch when the delivery system was pulled/retracted during the procedure.The physical evaluation of the device could not identify the conditions or circumstances that led to the pusher damage, but the damage is consistent with the device experiencing forces over specification.The returned detachment controllers were found to function as intended and would not have caused or contributed to the reported event.
|