SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101330-NLJ |
Device Problems
Melted (1385); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: b3: month and year of event have been provided, day is unknown.D4: udi section is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the included l-shaped connector had "melted".No adverse patient effects were reported by the customer.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
Other text: h3, h6 - updated one device and three photos were returned for evaluation.The photos received showed a damaged elbow.During visual inspection of the returned device the damaged elbow was detected as well, confirming the complaint.Based on the analysis conducted in the photos provided and the sample provided, the damaged elbow could had been generated by crushing, the most probable cause of this condition is improper handling of the product.No corrective actions are required.A review of the device of the device history records could not be completed as no lot number was provided by the customer.
|
|