Model Number 1458Q/86
Device Problems High impedance (1291); Failure to Sense (1559); Failure to Advance (2524); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, left ventricular lead failed to advance over the guidewire.The physician pushed the left ventricular lead to the target area until it could be placed.However, the lead exhibited failure to sense, high pacing impedance and high capture threshold.The lv lead was not used and replaced.There were no patient consequences.
 
Manufacturer Narrative
The reported events of failure to sense, unacceptable capture, high pacing impedance and failure to advance guidewire were not confirmed.Final analysis found that as received, a complete lead was returned in one piece.Visual and x-ray examination of the lead did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.A guidewire was able to be fully inserted and removed from the inner coil with no obstruction/restriction.S-curve was measured to be within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17368692
MDR Text Key319476250
Report Number2017865-2023-36399
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeTH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberS000089020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-