Model Number N ANTISERUM TO HUMAN ALBUMIN |
Device Problem
High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/25/2023 |
Event Type
malfunction
|
Event Description
|
A falsely elevated albumin result was obtained on a cerebrospinal fluid (csf) patient sample on a bn prospec system using n antiserum to human albumin reagent.The erroneous result was reported to the physician(s) and was questioned.On (b)(6) 2023, the sample was repeated on the same instrument for albumin csf, recovering lower.The lower result was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely elevated albumin result.
|
|
Manufacturer Narrative
|
An outside of the united states (ous) customer contacted a siemens customer care center (ccc).A siemens customer service engineer (cse) was dispatched to the customer¿s site.During this visit, the cse checked the probes, tubing, and syringes and did not detect any issues.The cse also cleaned a probe and the customer performed weekly maintenance.Furthermore, a system decontamination was performed.There is no evidence that a reagent or instrument contributed to the falsely elevated result.Quality controls recovered within acceptable ranges and no issues were reported with other patient samples.The calibration curve and the mean deviation were acceptable.A sample integrity issue contributed to the falsely elevated albumin result.The instrument and reagent are performing according to specifications.No further evaluation of this device is required.
|
|
Search Alerts/Recalls
|
|