Catalog Number 22442-19
Device Problems Difficult to Insert (1316); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2023
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a moderately tortuous, 90% stenosed lesion in the internal carotid artery.During preparation of an emboshield nav6 embolic protection system (eps), resistance was felt when pulling the filter into the delivery catheter (dc) pod with the torque device.The filter could not be fully pulled into the dc pod and became stuck.The filter was then difficult to remove from the dc pod.An unspecified stent was used to successfully complete the procedure.There was no patient involvement and no clinically significant delays in the procedure.No additional information was provided.
 
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Manufacturer Narrative
Visual, dimensional, and functional analysis was performed on the returned device.The reported difficulty to load the filter was unable to confirmed due to the condition of the returned delivery catheter.The reported difficult to remove was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no indication of a lot specific product quality issue.The investigation was unable to determine a conclusive cause for the reported difficulty.It may be possible that the filter was pulled into the pod too quickly or with excessive force causing the noted damage to the pod and resulting in difficulty pulling the filter back out due to the filter becoming stuck in the pod; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17371754
MDR Text Key319873935
Report Number2024168-2023-07826
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number22442-19
Device Lot Number2101061
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-