Model Number XS PST |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E3-occupation is patient/consumer.The investigation is ongoing.
|
|
Event Description
|
The customer complained of discrepant inr results with coaguchek xs meter serial number (b)(6) compared to a laboratory using an unknown reagent.The meter display has missing segments in the last number on the right side of the results area.Customer stated it looked like the right half of the number is visible and the left side is faded.There was no misinterpretation of results reported.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.The investigation found that the battery contacts as well as the circuit board are contaminated by liquid that has penetrated/corroded the solder contacts.This affects the battery contacts on the circuit board, which are interrupted by this and the conductive rubber contacts the cause of the issue is customer mishandling.Medwatch fields d9 and h3 have been updated.
|
|