It was reported that "after 3 minutes of use, the machine frequently alarms and the catheter may be kinked.Adjusting the equipment and catheter cannot solve the problem.After replacing the catheter, the problem was resolved".The 2nd catheter was inserted at the same insertion site.No patient harm or injury.The patient status is reported as "fine".
|
(b)(4).The reported complaint that the "machine frequently alarms and the catheter may be kinked" was confirmed based on the sample received.The customer returned a 40cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) with the original packaging box that matched the serial number on the returned sample (inp-2, inp-4) for investigation.The sample was returned in a cardboard box and was loosely packed within the original packaging box (inp-1, inp-5).Upon return, the peel away sheath was noted on the returned iabc (inp-6).The one-way valve was connected and tethered to the short driveline tubing (inp-7).The bladder was fully unwrapped (inp-8).A bend was noted to the iabc outer/central lumen at approximately 27.6cm from the iabc distal tip (inp-9).No other damage or abnormalities were noted to the returned sample.Dried blood was noted on the exterior surfaces of the returned iabc; no blood was noted within the helium pathway.The bladder thickness was measured at six points with measurements ranging from 0.0061in-0.0063in and was within specification of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute and then 30 seconds five separate times according to quality system document.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.No abnormalities or debris were noted.The iabc was leak tested using the lt-l-015 in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.Resistance was noted at approximately 27.8cm from the iabc distal tip, which is the location of the previously noted bend.The guidewire was able to advance through the central lumen.No blood or debris was noted.The guidewire was front loaded through the iabc luer.Resistance was noted at approximately 54.3cm from the iabc luer, which is the location of the previously noted bend.The guidewire was able to advance through the central lumen.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.Based on a review of the device history record (dhr), the product met specification upon release; however, the speci fications were not met during the complaint investigation due to the bent outer/central lumen.The root cause of the bent outer/central lumen was undetermined, but a most probable potential cause was due to customer handling.No further action was required at this time.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
|
It was reported that "after 3 minutes of use, the machine frequently alarms and the catheter may be kinked.Adjusting the equipment and catheter cannot solve the problem.After replacing the catheter, the problem was resolved".The 2nd catheter was inserted at the same insertion site.No patient harm or injury.The patient status is reported as "fine".
|