• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XACT CAROTID STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XACT CAROTID STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 82098-01
Device Problem Premature Activation (1484)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/21/2023
Event Type  malfunction  
Event Description
It was reported that the 10x30x136 xact stent system was removed packaging and noted the tip of the stent was exposed.Another xact stent was used to complete the procedure.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.Per device analysis the tip of stent was noted to be exposed; however, not flowered.The account confirmed this is correct.
 
Manufacturer Narrative
The device was returned for analysis.The reported premature activation was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that inadvertent mishandling/inadvertently rotating the deployment handle clockwise while unpackaging the device caused the sheath to slightly retract thus resulting in the stent to be partially exposed resulting in the reported premature activation.The noted sheath kink likely occurred due to handling during packing for return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Type of Device
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17384064
MDR Text Key319625441
Report Number2024168-2023-07915
Device Sequence Number1
Product Code NIM
UDI-Device Identifier08717648010347
UDI-Public08717648010347
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P040038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number82098-01
Device Lot Number2101061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-