Concomitant medical products: 694758 lead <(>&<)> 5076-45 lead implanted: (b)(6) 2011 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
It was reported that the patient had exertional fatigue and shortness of breath.It was found that the right ventricular (rv) lead had a lead integrity alert (lia) triggered by sensing integrity counter (sic), and rv pacing impedance variation.It was further noted that there was rv oversensing.It was also reported that the episodes counters were off on the reports on the network.It was further reported from follow up with the clinic that the patient had been using electrical tools which caused a short period of electromagnetic interference (emi) with the cardiac resynchronization therapy defibrillator (crt-d).The emi on the device resulted in the ventricular oversensing and triggered the lia for sic and pacing impedance variation that was observed on the rv channel.Follow up with the clinic indicated that no intervention was needed, and the device remains in use.No further patient complications have been reported as a result of this event.
|