Model Number 71992-01
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Pain (1994); Device Embedded In Tissue or Plaque (3165)
Event Date 07/07/2023
Event Type  Injury  
Manufacturer Narrative
The reported product is not expected to be returned, as the device was reportedly discarded.A follow-up report will be submitted if product is returned or additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported, that the adc device sensor filament remained in skin after sensor removal.The customer reported, experiencing pain.And had contact with a healthcare provider for "removal through minor surgery".There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned and a valid serial number has not been provided.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that freestyle libre sensor continue to be safe, effective, and perform as intended in the field.A tripped trend review was completed for the reported complaint and fs libre sensors, and there were no adverse trends that indicate any potential product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported that the adc device sensor filament remained in skin after sensor removal.The customer reported experiencing pain, and had contact with a healthcare provider for "removal through minor surgery".There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key17386309
MDR Text Key319650736
Report Number2954323-2023-31963
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? No
Date Manufacturer Received08/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-