• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2023
Event Type  malfunction  
Event Description
During an implant procedure, the left ventricular (lv) was unable to separate from the guidewire resulting in lead insulation damage.The lv lead was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported events of stylet could not be removed and insulation damage were confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found that the connector pin and crimp sleeve were pulled out of the connector assembly along with the inner coil which is consistent with procedural damage and the ptfe stylet coating was bunched up/clogged with the inner coil distal to the connector pin.The cause of the reported event was due to bunched up ptfe coating of the stylet inside the inner coil that prevented the removal of the stylet and excessive forces resulted in the connector pin and crimp sleeve to be pulled out of the connector assembly.Further analysis found procedural damage to the lead insulation distal to the connector boot.The cause of the reported event of insulation damage was isolated to procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17392756
MDR Text Key319788735
Report Number2017865-2023-36700
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000090630
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-