U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNK HIP ACETABULAR CUP PINNACLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNK HIP ACETABULAR CUP PINNACLE Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP PINNACL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Necrosis (1971); Pain (1994); Discomfort (2330); Metal Related Pathology (4530); Swelling/ Edema (4577)
Event Date 02/14/2023
Event Type  Injury  
Event Description
Metal on metal medical records received.After review of the medical records the patient was revised to address metallosis and extensive abductor muscle necrosis resulting to right hip pain, elevated metal ions, corrosion, pseudotumor formation and discomfort.Operative note reported adipose tissue and bursal tissue were dissected.There was a fluid in the joint under tension and had evidence of metal debris and abductor muscle and capsular tissue necrosis.This was carefully debrided.Patient had a small amount of shuck tolerated hip extension and external rotation to about 70 degrees.Minimal corrosion in taper stem and femoral head.Removed necrotic tissues around the acetabulum.There was extensive corrosion at the taper of the liner component surface and with pseudomembrane formation on the undersurface of the acetabular liner.Cup and stem are well fixed.Doi: (b)(6) 2011; dor: (b)(6) 2023 ; right hip.
 
Manufacturer Narrative
Product complaint #: (b)(4).D4-the device catalog number is unknown; therefore, udi is unavailable.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK HIP ACETABULAR CUP PINNACLE
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17396413
MDR Text Key319832833
Report Number1818910-2023-15211
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR CUP PINNACL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received07/26/2023
Supplement Dates Manufacturer Received08/17/2023
Supplement Dates FDA Received08/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR LINER METAL ULTIMA.; UNK HIP FEMORAL HEAD METAL.; UNK HIP FEMORAL SLEEVE SROM.; UNK HIP FEMORAL STEM SROM.
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexMale
-
-