MAUDE Adverse Event Report: ZIMMER BIOMET, INC. HEADLESS TROCAR DRILL PIN 3.2 MM DIAMETER 75 MM LENGTH; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. HEADLESS TROCAR DRILL PIN 3.2 MM DIAMETER 75 MM LENGTH; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Event Description
It was reported that the pin got stuck in a tibial cut guide.The pin was returned and determined to be fractured.The rep believes it fractured when the surgeon was putting the pin into the guide.There was no retained hardware left in the patient and there was no consequences or impact to the patient.
 
Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.Visual examination of the returned product identified the pin is seized in a guide hole of the cut guide and the pin is fractured.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADLESS TROCAR DRILL PIN 3.2 MM DIAMETER 75 MM LENGTH
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17397541
MDR Text Key319792725
Report Number0001822565-2023-02003
Device Sequence Number1
Product Code LXH
UDI-Device Identifier00889024201620
UDI-Public(01)00889024201620(17)330324(10)65711440
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00590102000
Device Lot Number65711440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/05/2023
Initial Date FDA Received07/26/2023
Supplement Dates Manufacturer Received10/24/2023
Supplement Dates FDA Received10/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-