Model Number N/A |
Device Problems
Fracture (1260); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the pin got stuck in a tibial cut guide.The pin was returned and determined to be fractured.The rep believes it fractured when the surgeon was putting the pin into the guide.There was no retained hardware left in the patient and there was no consequences or impact to the patient.
|
|
Manufacturer Narrative
|
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.Visual examination of the returned product identified the pin is seized in a guide hole of the cut guide and the pin is fractured.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|