Manufacturer's investigation conclusion: the centrimag console, serial number (b)(6), was evaluated due to the reported event of the centrimag motor, serial number (b)(6), not functioning as intended.The centrimag console was not returned for analysis, and no log files were associated with the reported event.The investigation of the returned centrimag motor confirmed that the cause of the reported event was due to an issue with the motor and has been addressed via the motor¿s investigation.Review of the device history record for the centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 3 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 9 entitled "emergency / troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." no further information was provided.The manufacturer is closing the file on this event.
|