Model Number 71992-01 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problems
Headache (1880); Hypoglycemia (1912); Vomiting (2144); Discomfort (2330)
|
Event Date 07/12/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the abbott diabetes care (adc) device.Customer received a "scan again in 10 minutes" message and was unable to obtain readings.As a result, customer experienced "felt ill", "headaches", "vomiting" and was unable to self-treat, requiring third-party treatment of "4 sugar" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed on the sensor patch.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing and poise voltage testing (simulation of the electrical signal produced by the sensor tail) was performed and all results were within specification.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the abbott diabetes care (adc) device.Customer received a "scan again in 10 minutes" message and was unable to obtain readings.As a result, customer experienced "felt ill", "headaches", "vomiting" and was unable to self-treat, requiring third-party treatment of "4 sugar" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
|