• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510880
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2023
Event Type  Injury  
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) stone removal procedure performed on (b)(6) 2023.During the procedure, a trapezoid rx basket was used in an attempt to crush a 1 cm stone.However, the handle broke and basket was unable to fracture the stone.Moreover, the tip of the basket did not deploy as intended, causing the basket to lodge in the patient's bile duct.A cre balloon was used to dilate the duct and ampulla, allowing the scope with the basket containing the stone to be remove from the bile duct.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the device expiration and manufacture dates are unknown.Block h6: imdrf device code a0401 captures the reportable event of handle break.Imdrf device cod a150301 captures the reportable event of tip failure to separate.Imdrf code f2301 captures the event of cre balloon was used to dilate the duct and remove the scope along with the basket containing the stone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17405898
MDR Text Key319858983
Report Number3005099803-2023-03979
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296393
UDI-Public08714729296393
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00510880
Device Catalogue Number1088
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-